Vitatron C20 SR, Vitatron C10 S

Programming Parameters

Pacing and Sensing Parameters

Parameter Name Range Delivery Settings Per Model
Vitatron C20 SR
C20A1/C20A2
Vitatron C10 S
C10A1/C10A2
Mode VVIR, VVI, VVT, VOO, AAIR, AAI, AAT, AOO, OOO VVI VVI
Lower rate 40-(5)-130 min-1 60 min-1 60 min-1
Maximum pacing rate 90-(5)-130 min-1 120 min-1
90-(5)-170 min-1 120 min-1
Pulse amplitudea 0.5-(0.25)-4.0-(0.5)-8.0 V 3.75 V 3.75 V
Pulse durationa 0.1-(0.05)-1.0 ms 0.4 ms 0.4 ms
Atrial sensitivity Uni: 0.5-(0.1)-1.0-(0.5)-7.5 mV
Bi: 0.25, 0.3-(0.1)-1.0- (0.5)-7.5 mV
0.7 mV 0.7 mV
Ventricular sensitivity 1.0-(0.5)-10.0 mV 2.0 mV 2.0 mV
Sensing/pacing polaritya Unipolar, Bipolar Unipolar Unipolar
Refractory period 250-(10)-500 ms 330 ms 330 ms
Conditional hysteresis 0-(5)-30 min-1 0 min-1 0 min-1

a Conditional hysteresis is only applicable in VVI, VVT, AAI and AAT modes.

Other Programmable Parameters

Parameter Name Range Delivery Settings Per Model
Vitatron C20 SR
C20A1/C20A2
Vitatron C10 S
C10A1/C10A2
Flywheel On, Off Off Off
Night lower rate 40-(5)-130 min-1 60 min-1 60 min-1
Start of night 18:00-(5 min)-02:55 hh:mm 00:00 00:00
End of night 04:00-(5 min)-11:55 hh:mm 06:00 06:00
Pacemaker time 00:00-(1 min)-23:59 hh:mm 00:00 00:00
Slope Auto, Fixed Auto
Threshold Low, Low/Med, Medium, Med/High, High Medium
Acceleration Fast, Standard, Gradual Standard
Deceleration Fast, Standard, Gradual Standard
Mode switching Auto, Fixed Fixed Fixed
Ventricular Rate Stabilization On, Off Off (C20A2 only)
Maximum therapy rate 70-(10)-120 min-1 100 min-1 (C20A2 only)

Tolerances (valid between 22 °C and 45 °C throughout the pacemaker lifetime)

Parameter Tolerance
Lower rate (min-1) Programmed value ±15 ms
Maximum pacing rate (min-1) Programmed value ±15 ms
Maximum tracking rate (min-1) Programmed value ±15 ms
Pulse amplitude (V) Programmed value ≤ 1.0: +40%/-10%
Programmed value > 1.0 ≤ 4.0: +20%/-10%
Programmed value > 4.0 ≤ 7.0: +10%/-10%
Programmed value > 7.0: +10%/- 20%
Pulse duration (ms) Programmed value - 0.02/+0.04
Atrial sensitivity (mV) Programmed value ±(10% +0.16) at 37 °Ca
Ventricular sensitivity (mV) Programmed value ±(20% +0.16) at 37 °Ca
Refractory period (ms) Programmed value ±15
Hysteresis (conditional)-1 Programmed value ±15
Lead impedance Measured value ±(20% +20)
Battery voltage (V) Measured value ±0.03 V
EGM amplitude (atrial) ±(10% +0.16 mV) +1% of full scale (at 37 °C)a
EGM amplitude (ventricular) ±(20% +0.16 mV) +1% of full scale (at 37 °C)a
ECG marker timing ±10 ms

a Add ±10% over temperature range 22 °C to 45 °C.

Telemetered patient and pacemaker data

Data item
Patient’s name and identification code
Patient’s date of birth
Pacemaker type, model number and serial number
Pacemaker implantation date
Lead manufacturer, model number and serial number
Lead implantation date
Pacemaker dependent (yes/no)
Symptoms
Indications for pacing
Etiology
Pacemaker time
Notes
Physician’s name and phone number

Measurements

Measurement
Remaining pacemaker longevity
Lead polarity (atrial and ventricular)
Lead impedance (atrial and ventricular)
Pulse amplitude (atrial and ventricular)
Battery status
Remaining battery capacity
Battery impedance
Battery voltage
Mean battery current
Consumed battery charge
Pulse energy
Average pulse current
P/R-wave amplitude
Stimulation threshold
Fast learning (C20A1 and C20A2 only)

Diagnostic Capabilities

Category Feature
ECG and therapy analysis Intracardiac ECG
ECG event marker intervals
ECG marker annotation
Therapy Advisor
Holters 24-hour Holter
30-minute Holter
Histograms/tables Diurnal rhythm distribution
P-wave amplitude (AAI and AAIR modes)
Ventricular rate (VXX modes)
Ventricular rate irregularity (VXX modes, C20A2 only)
Atrial rate (AAI, AAIR and AAT modes)
History(CXXA2 only) Stimulation threshold
Lead impedance
P-wave amplitude
Percentage paced
Counters Percentage of atrium or ventricle paced
Percentage of ventricle sensed (VXX modes)
Percentage of physiologic atrial senses (AAI, AAIR and
AAT modes)
Percentage of pathologic atrial senses (AAI, AAIR and AAT
modes)
Number of episodes of atrial tachyarrhythmia (AAI, AAIR
and AAT modes)
Number of PACs (AAI, AAIR and AAT modes)
Number of accelerometer counts (AAIR and VVIR

Technical Parameters

Magnet Mode

Parameter Rate
Magnet pacing mode AOO or VOO
Magnet rate, battery status "Good" 100 min-1 (600 ms)
Magnet rate, battery status "Ageing" 95 min-1 (630 ms)
Magnet rate, battery status "Replace PM" 86 min-1 (700 ms)

Warning: In magnet mode the pacemaker operates in an asynchronous pacing mode. If the intrinsic rate is higher than the magnet rate, this may induce ventricular tachycardia or ventricular fibrillation.

Exposed materials

Item Materials
Can Titanium
Connector Polyether-Urethane (PUR) and silicone rubber

Physical characteristics

Characteristic Value
Dimensionsa 46.8 x 45.9 x 7.25 mm
Mass 27.3 ± 0.5 g
Volume 11.9 ± 0.5 cm3
Surface area 33.1 cm2
X-ray identification VF
Connector IS-1 dual chamber

a Connector thickness may vary.

Electrical specifications

Characteristic Value
Output capacitor 3.4 μF
Input impedance ≥ 100 kΩ
Current/Energy consumption VVIR VVI
Pacinga 16.2 μA 15.3 μA
Inhibited 10.3 μA 9.3 μA

a 100% pacing at 60 min-1, 3.75 V, 0.4 ms, 500 Ω.

Power source

Characteristic Value
Cell type Pi 223 lithium iodine
Voltage 2.8 V (RRTa 2.6 V)
Capacity 1.4 Ah (RRTa 0.1 Ah)
Battery manufacturer MECC

a RRT = Recommended Replacement Time.

Interference and high rate protection

Characteristic Value
Interference detection frequency 15.4 ±0.5 Hz (923 ±29 min-1)
Pacing mode during interference VOO(R) if programmed to a ventricular mode.
AOO(R) if programmed to an atrial mode.
OOO if programmed to the OOO mode
Pacing rate during interference Pacemaker behaves as if no spontaneous cardiac activity
is present. If interference continues, the eventual pacing
rate is either the sensor-indicated rate or the lower rate,
whichever is the higher.
Ventricular high rate protection 205 ±10 min-1

a Atrial high rate protection for pacemakers programmed to AXX modes; ventricular high rate protection for pacemakers programmed to VXX modes.