Vitatron C70 DR, C60 DR, Vitatron C50 D

Programming Parameters

Pacing and Sensing Parameters

Parameter Name Range Delivery Settings Per Model
Vitatron C70 DR
C70A3
Vitatron C60 DR
C60A1/C60A2
Vitatron C50 D
C50A1/C50A2
Mode DDDR, DDIR, DDD, DDI, DOO, VDDR, VDD, VVIR, VVI, VVT, VOO, AAIR, AAI,
AAT, AOO, OOO
DDD DDD DDD
Lower rate 40-(5)-130 min-1 60 min-1 60 min-1 60 min-1
Maximum pacing rate 90-(5)-170 min-1 120 min-1 120 min-1 120 min-1
Maximum tracking rate 90-(5)-190 min-1 140 min-1 140 min-1 140 min-1
Pulse amplitudea 0.5-(0.25)-4.0-(0.5)-8.0 V 3.75 V 3.75 V 3.75 V
Pulse durationa 0.1-(0.05)-1.0 ms 0.4 ms 0.4 ms 0.4 ms
Atrial sensitivity Uni: 0.5-(0.1)-1.0-(0.5)-7.5 mV
Bi: 0.25, 0.3-(0.1)-1.0- (0.5)-7.5 mV
0.7 mV 0.7 mV 0.7 mV
Ventricular sensitivity 1.0-(0.5)-10.0 mV 2.0 mV 2.0 mV 2.0 mV
Sensing/pacing polaritya Unipolar, Bipolar Unipolar Unipolar Unipolar
Atrial blanking on VP 50-(25)-300 ms 150 ms 150 ms 150 ms
Atrial blanking on VS 25-(25)-150 ms 50 ms 50 ms 50 ms
Ventricular blanking on AP 20-(5)-50 ms 30 ms 30 ms 30 ms
Ventricular safety pacing On, Off On On On
Atrial refractory period 250-(10)-500 ms 330 ms 330 ms 330 ms
Ventricular refractory period 250-(10)-500 ms 260 ms 260 ms 260 ms
Maximum sensed AV delay 45-(5)-260 ms 120 ms 120 ms 120 ms
Maximum paced AV delay 80-(5)-300 ms 160 ms 160 ms 160 ms
Sensed/Paced AV offset 20-(5)-50 ms 40 ms 40 ms 40 ms
Adaptive AV delay Off, Medianb, Fastc Median Median Median
Refined Ventricular Pacing On, Off Off Off Off
AV delay extension 60-(20)-120 ms 60 ms 60 ms 60 ms
Refined Atrial Pacing On, Off Off Off Off
Atrial hysteresis 100-(25)-200 ms 100 ms 100 ms 100 ms

a Independently programmable in atrium and ventricle.
b Median = 5 ms per atrial rate change of 10 min-1.
c Fast = 10 ms per atrial rate change of 10 min-1.

Other Programmable Parameters

Parameter Name Range Delivery Settings Per Model
Vitatron C70 DR
C70A3
Vitatron C60 DR
C60A1/C60A2
Vitatron C50 D
C50A1/C50A2
Flywheel On, Off Off Off Off
Tachy fallback rate Off, 45-(5)-100 min-1 Off Off
Night tachy fallback rate 40-(5)-100 min-1 60 min-1 60 min-1
Night lower rate 40-(5)-130 min-1 60 min-1 60 min-1
Start of night 18:00-(5 min)-02:55 hh:mm 00:00 00:00 00:00
End of night 04:00-(5 min)-11:55 hh:mm 06:00 06:00 06:00
Pacemaker time 00:00-(1 min)-23:59 hh:mm 00:00 00:00 00:00
Post-PVC response On, Off On On On
PVC synchronous Astim On, Off Off Off Off
Sensor Slope Auto, Fixed Auto Auto Auto
Activity Threshold Low, Low/Med, Medium, Med/High, High Medium Medium Medium
Activity Acceleration Fast, Standard, Gradual Standard Standard Standard
Activity Deceleration Fast, Standard, Gradual Standard Standard Standard
Mode switching Auto, Fixed Fixed Fixed Fixed
Mode switching sensitivity Standard, Moderate Standard Standard Standard
ASP interval 250-(5)-400 ms 300 ms 300 ms 300 ms
A burst ratea 100-(10)-320- (15)-425 min-1 130 min-1 130 min-1 130 min-1
VOO backupa On, Off Off Off Off
Ventricular Rate Stabilization On, Off Off Off Off
Triggered overdrive pacingb On, Off Off
PAC Suppression On, Off Off
Post-PAC Response On, Off Off
Maximum therapy rate 70-(10)-120 min-1 100 min-1 100 min-1 100 min-1
Selected Episodes trigger Off, Atrial Rate, Ventricular Rate Off Off Off
EGM recording On, Off Off Off Off

a Only temporarily programmable.
b When triggered overdrive is programmed to "On", both triggered overdrive pacing therapies (PAC Suppression and Post-PAC Response) are programmed on.

Tolerances (valid between 22 °C and 45 °C throughout the pacemaker lifetime)

Parameter Tolerance
Lower rate (min-1) Programmed value ±15 ms
Maximum pacing rate (min-1) Programmed value ±15 ms
Maximum tracking rate (min-1) Programmed value ±15 ms
Pulse amplitude (V) Programmed value ≤ 1.0: +40%/-10%
Programmed value > 1.0 ≤ 4.0: +20%/-10%
Programmed value > 4.0 ≤ 7.0: +10%/-10%
Programmed value > 7.0: +10%/- 20%
Pulse duration (ms) Programmed value - 0.02/+0.04
Atrial sensitivity (mV) Programmed value ±(10% +0.16) at 37 °Ca
Ventricular sensitivity (mV) Programmed value ±(20% +0.16) at 37 °Ca
Blanking period (ms) Programmed value ±15
Refractory period (ms) Programmed value ±15
Sensed/paced AV delay (ms) Programmed value ±15
Sensed/paced AV offset (ms) Programmed value ±15
Atrial hysteresis (ms) Programmed value ±15
Lead impedance (Ω) Measured value ±(20% +20)
Battery voltage (V) Measured value ±0.03
EGM amplitude (atrial) ±(10% +0.16 mV) +1% of full scale (at 37 °C)a
EGM amplitude (ventricular) ±(20% +0.16 mV) +1% of full scale (at 37 °C)a
ECG marker timing ±10 ms

a Add ±10% over temperature range 22 °C to 45 °C.

Telemetered patient and pacemaker data

Data item
Patient’s name and identification code
Patient’s date of birth
Pacemaker type, model number and serial number
Pacemaker implant date
Lead manufacturer, model number and serial number (atrial and ventricular)
Lead implant date (atrial and ventricular)
Pacemaker dependent (yes/no)
Symptoms
Indications for pacing
Etiology
Pacemaker time
Anticoagulation is applied (yes/no)
Notes
Physician’s name and phone number

Measurements

Measurement
Remaining pacemaker longevity
Lead polarity (atrial and ventricular)
Lead impedance (atrial and ventricular)
Pulse amplitude (atrial and ventricular)
Battery status
Remaining battery capacity
Battery impedance
Battery voltage
Mean battery current
Consumed battery charge
Average pulse current
Pulse energy
P-wave amplitude
R-wave amplitude
Stimulation threshold
VA interval
Fast Learning

Diagnostic Capabilities

Category Feature
ECG and therapy analysis Intracardiac ECG
ECG event marker intervals
ECG marker annotation
Therapy Advisor
Holters 24-hour Holter
30-minute Holter
Histograms/tables Diurnal rhythm distribution
P-wave amplitude
Atrial rate
Ventricular rate
Ventricular rate irregularity
VA interval
History(CXXA2 only) Stimulation threshold (A and V)
Lead impedance (A and V)
P-wave amplitude
Percentage paced (A and V)
Percentage of atrial rate > 200 min-1
Percentage AV synchrony
Counters Percentage of atrium paced
Percentage of physiologic atrial senses
Percentage of pathologic atrial senses
Percentage of atrial rate > 200 min-1 (CXXA2 only)
Percentage of ventricle paced
Percentage of ventricle sensed
Percentage of AV synchrony
Mean V rate during A tachy
Number of episodes of atrial tachyarrhtythmia
Number of episodes of retrograde conduction
Number of PACs
Number of PVCs
Number of VSPs
Number of accelerometer counts

Technical Parameters

Magnet Mode

Parameter Rate
Magnet pacing mode Fixed rate pacing in programmed mode.
Magnet rate, battery status "Good" 100 min-1 (600 ms)
Magnet rate, battery status "Ageing" 95 min-1 (630 ms)
Magnet rate, battery status "Replace PM" 86 min-1 (700 ms)

Warning: In magnet mode the pacemaker operates in an asynchronous pacing mode. If the intrinsic rate is higher than the magnet rate, this may induce ventricular tachycardia or ventricular fibrillation.

Exposed materials

Item Materials
Can Titanium
Connector Polyether-Urethane (PUR) and silicone rubber

Physical characteristics

Characteristic Value
Dimensionsa 50.9 x 45.9 x 7.25 mm
Mass 28.6 ± 0.5 g
Volume 12.7 ± 0.5 cm3
Surface area 33.1 cm2
X-ray identification VF
Connector IS-1 dual chamber

a Connector thickness may vary.

Electrical specifications

Characteristic Value
Output capacitor 3.4 μF
Input impedance: atrium ≥ 100 kΩ
Input impedance: ventricle ≥ 100 kΩ
Current/Energy consumption DDDR DDD VVIR VVI
Pacinga 24.2 μA 23.7 μA 15.4 μA 14.8 μA
Inhibited 12.9 μA 12.3 μA 9.4 μA 8.9 μA

a 100% pacing at 60 min-1, 3.75 V, 0.4 ms, 500 Ω.

Power source

Characteristic Value
Cell type Pi 223 lithium iodine
Voltage 2.8 V (RRTa 2.6 V)
Capacity 1.4 Ah (RRTa 0.1 Ah)
Battery manufacturer MECC

a RRT = Recommended Replacement Time.

Interference and high rate protection

Characteristic Value
Interference detection frequency 15.4 ±0.5 Hz (923 ±29 min-1)
Pacing mode during interference DOO(R) if programmed to a dual chamber mode.
VOO(R) if programmed to a ventricular mode.
AOO(R) if programmed to an atrial mode.
OOO if programmed to the OOO mode.
Pacing rate during interference Pacemaker behaves as if no spontaneous cardiac activity
is present. If interference continues, the eventual pacing
rate is either the sensor-indicated rate or the lower rate,
whichever is the higher.
Ventricular high rate protection 205 ±10 min-1